Polarographic Determination of Lead
in Lead Plating Solutions
Rafael Diaz, Minneapolis-Honeywell
Regulator Company, Minneapolis, MN
A RAPID CONTROL method employing
the use of the polarograph is described for the analysis of lead in lead plating
solutions. The diffusion current of the lead ion in a medium containing potassium
nitrate as indifferent electrolyte is very well defined and forms the basis
for this determination. The time required for an analysis is just a fraction
of that required when lead is analyzed by the common acid-sulfate gravimetric
procedure because no weighing of the sample, precipitation or drying to constant
weight are needed in the procedure.
According to Kolthoff and Linganet,
the reduction of lead ion is reversible at the dropping mercury electrode and
the diffusion current is very well defined in the presence of a small amount
of gelatin. In 1N potassium nitrate the half-wave potential is found to be 0.703
volt when a stationary pool of mercury is used as the anode. This potential
is quite reproducible under the experimental conditions. The concentration of
gelatin, while not critical, should be kept below 0.01 per cent and the amount
added provides for a final concentration of gelatin of about 0.004 per cent.
This may be substituted by an even smaller concentration of the sodium salt
of methyl red (about 2 x 10 per cent). Methyl red is itself reduced at the dropping
electrode, but the concentration used is so small that it will have no appreciable
effect on the diffusion current of the lead ion.
Fig. 1--Polarogram
of lead in lead plating solution.
Fig.2--Calibration
curve for lead in lead plating solutions. Data from Table 1.
The author knows of no interferences
in the procedure with the exception of tin, whose chlorostannic complex gives
a wave at0.48 volt (vs. a stationary mercury pool). This interference
would only be serious if tin were present with enough chloride to form a chlorostannic
complex and in such high concentration that its wave would obscure that of the
lead, which appears at a more negative potential. If such were the case, a method
of separation would have to be used. Kolthoff and Linganet make use of sodium
hydroxide for the simultaneous analysis of tin and lead and obtain two widely
separated waves, thus eliminating any possibility of obtaining the two waves
at approximately the same potential. This method seems very suitable for application
in the case of the tin-lead alloys.
Since tin is not encountered in lead plating solutions except possibly as an
impurity it would not interfere with the lead determination because its concentration
would be so small that no danger of obtaining a coalescing wave with tin would
be present.
A typical polarogram of lead in a plating solution is shown in Fig. 1.
TABLE
I. DATA FROM POLAROGRAMS OF LEAD PLATING SOLUTIONS OF VARIOUS CONCENTRATIONS*
Conc.
of Lead, oz/gal
Diffusion
in current in microamperes**
Diffusion
current, Microamps per oz/gal**
0
...
...
4
1.09
0.273
8
2.15
0.269
12
3.25
0.271
16
4.32
0.270
20
5.39
0.270
24
6.47
0.270
28
7.57
0.270
Avg.
0.270 ± 0.007
*Lead plating
solutions of various concentrations of lead ion in 2M potassium nitrate
and gelatin as directed by the procedure in the text. Air removed from the
solutions with nitrogen. t = 3.80 sec., m2/3t1/6
= 2.30 mg2/3sec-1/2. Diffusion
currents measured at a potential of -0.703 volt with respect to a stationary
mercury pool.
**Corrected.
REAGENTS AND APPARATUS
2M Potassium nitrate solution
Gelatin 1 per cent solution
Mercury, triple distilled, for the mercury pool and the dropping electrode
Sargent manual polarograph, Model 3
PROCEDURE Calibration
A calibration chart was constructed by measuring the diffusion current of several
lead plating solutions prepared in the laboratory from analytical reagents This
calibration chart is, of course, valid only for the particular instrument and
electrode with which it was prepared; if for any reason it becomes necessary
to replace the capillary, then the curve must be redetermined. Even if no such
replacement is necessary, it is advisable to make regular determinations on
standard solutions to determine whether the characteristics of the instrument
have varied with time. This is especially important as regards the sensitivity
of the galvanometer. The data obtained from these polarograms, and the diffusion
current constants calculated therefrom, are shown in Table I. The data from
Table I are shown plotted in graph form in Fig. 2.
Lead Plating Solution
Pipette 1.0 ml of sample into a 500 ml volumetric flask. Add 4 ml of a 1 per
cent solution of gelatin and dilute with water to the mark. Pipette 5.0 ml of
this solution into the polarographic cell and add 5.0 ml of 2M potassium nitrate.
Record the polarogram between0.50 and 1.00 volts, applying a bridge
potential of 1.0 volt.
TABLE
II. RESULTS OF ANALYSES OF LEAD SOLUTIONS*
Solution No.
Amount Added
Volumetric Method
Polarographic Method
1
27.2
27.1
27.1
2
25.7
25.6
25.4
3
20.8
20.7
20.9
4
25.0
24.8
24.9
5
23.5
23.5
23.3
*Concentration
of lead expressed in oz/gal.
RESULTS AND DISCUSSION
Table II shows the agreement obtained between the polarographic and gravimetric
methods for the analysis of lead in some plating solutions. The figures show
that the method gives results in agreement with the gravimetric values to within
less than 2 per cent in the range of concentrations studied. Such accuracy is
perfectly adequate for the purpose of routine control of the lead plating baths.
LITERATURE CITED
Kolthoff and Lingane, Polarography, Interscience Publishers, Inc.
The information contained in this site is provided for your review and convenience. It is not intended to provide legal advice with respect to any federal, state, or local regulation.
You should consult with legal counsel and appropriate authorities before interpreting any regulations or undertaking any specific course of action.
Please note that many of the regulatory discussions on STERC refer to federal regulations. In many cases, states or local governments have promulgated relevant rules and standards
that are different and/or more stringent than the federal regulations. Therefore, to assure full compliance, you should investigate and comply with all applicable federal, state and local regulations.
The information contained in this site is provided for your review and convenience. It is not intended to provide legal advice with respect to any federal, state, or local regulation.
You should consult with legal counsel and appropriate authorities before interpreting any regulations or undertaking any specific course of action.
Please note that many of the regulatory discussions on STERC refer to federal regulations. In many cases, states or local governments have promulgated relevant rules and standards
that are different and/or more stringent than the federal regulations. Therefore, to assure full compliance, you should investigate and comply with all applicable federal, state and local regulations.